4 • Troubleshooting
Doc. 065378-01 1/12
81
procedures. If the detector fails the wavelength calibration again, contact
Thermo Fisher Scientific for assistance.
•
Deuterium lamp intensity out of calibration
1. Remove the cell from the optical bench (see
2. Check the deuterium lamp intensity:
a. Open the Chromeleon 7 Command window (see
) or the
Chromeleon 6.8 Commands dialog box (see
).
b. Select
UV
, and then select
LampIntensity
.
If the lamp intensity reading is above 10 million counts for a standard or
semi-preparative cell
or
above 3 million counts for a semi-micro cell,
follow these steps:
a. Open the Chromeleon 7 Command window (see
) or the
Chromeleon 6.8 Commands dialog box (see
).
b. Select
UV
, and then select
UV_Calibration
.
c. Execute the
Intensity
command to calibrate the lamp intensity.
d. Execute the
Autozero
command to run the autozero routine.
e. If no error message is displayed, flush the cell with the mobile
phase for the application and/or clean the cell as instructed in
.
f. Reinstall the cell in the optical bench.
g. Autozero the detector again to verify that the problem has been
resolved. If the error message appears again, contact Thermo
Fisher Scientific for assistance.
If the lamp intensity reading is below 10 million counts for a standard or
semi-preparative cell
or
below 3 million counts for a semi-micro cell,
follow these steps:
a. Check that the flange of the deuterium lamp is aligned with the
two locating pins in the optical bench (see
). If the
flange is aligned, the lamp is defective and should be replaced
(see
b. If the error message appears again, contact Thermo Fisher
Scientific for assistance.
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...