Dionex PDA Operator’s Manual
76
Doc. 065378-01 1/12
•
Connect a backpressure line to the cell (see
). This
increases the backpressure on the cell, thereby shrinking bubbles and
allowing them to pass more easily through the cell.
•
Contaminants in flow cell
Clean the cell (see
•
Incorrect lamp selected
If a UV wavelength is selected, make sure the deuterium lamp is on. If a
visible wavelength is selected, make sure the tungsten lamp is on. When a
lamp is on, the corresponding LED (
DEUTERIUM – UV
or
TUNGSTEN – VIS
) is
lighted.
•
Lamp needs replacement
The light output of the lamp decreases over time. The lower light output may
increase noise above an acceptable level, especially if the lamp has been in
service for more than 2000 hours. If this occurs, replace the lamp. See
or
for instructions on replacing the deuterium lamp or
tungsten lamp, respectively.
•
Lamp incorrectly installed
Make sure that both lamps are fully seated in the optical bench. See
for installation instructions for the deuterium lamp
or tungsten lamp, respectively.
•
Inappropriate reference wavelength
Select a reference wavelength in a quiet area of the spectrum where little
absorption occurs. Thermo Fisher Scientific recommends selecting a
reference wavelength below 380 nm for UV applications and above 600 nm
for visible applications. For more information, see
.
•
Bandwidth is too narrow
Select a wider bandwidth and reference bandwidth, if appropriate for the
application. For guidelines when selecting the sample and reference
bandwidths, see
and
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...