Dionex PDA Operator’s Manual
108
Doc. 065378-01 1/12
A.5.3
Semi-Preparative Flow Cell
A.6 Heat Exchangers
Chemical
Compatibility
•
Do not use bases stronger than 0.1 M; these solutions will
etch the fused silica windows of the flow cell.
•
Do not use normal phase or chlorinated solvents with the
PEEK flow cell; these solutions will damage the cell.
•
Do not use low pH solutions with the stainless steel flow
cell; these solutions may cause corrosion, contamination,
and metal leaching.
Cell Body
PEEK
Volume and
Optical Path
Length
0.7
P
L volume; 0.4 mm (0.02 in) path length
Maximum
Operating
Pressure
10.34 MPa (1500 psi)
Chemical
Compatibility
•
Do not use bases stronger than 0.1 M; these solutions will
etch the fused silica windows of the flow cell.
•
Do not use normal phase or chlorinated solvents with the
flow cell; these solutions will damage the cell.
Maximum
Operating
Pressure
PEEK cell: 2 MPa (300 psi)
Stainless steel cell: 3 MPa (500 psi)
Volume of
Standard Cell
Heat Exchangers
(including cell
inlet tubing)
PEEK cell: 45
P
L
Stainless steel cell: 20
P
L
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...