B • Installation
B.1 Facilities Requirements
•
Make sure the installation site for the Thermo Scientific Dionex ICS Series
Photodiode Array Detector (Dionex PDA) meets the power and
environmental specifications listed in
•
For mobile phase generation, or when manually preparing mobile phase and
regenerant, use ASTM Type I (18 megohm-cm) filtered and deionized water
that meets the specifications listed in
.
•
Optical detectors are sensitive to vibration. Provide a sturdy, vibration-free
workbench of a height that allows convenient access to the interior of the
Dionex PDA.
•
Allow at least 6 cm (2.4 in) behind the Dionex PDA for power connections
and ventilation.
Do not block air flow to the detector.
During operation,
room air must be inducted into the detector to cool the electronics and
dissipate the heat generated by the lamps.
Contaminant
Specification
Ions–Resistivity
>18.0 (megohm-cm)
Organics–TOC
<10 ppb
Iron/Transition Metals*
<1 ppb
Pyrogens
<0.03 (Eu/mL)
Particulates > 0.2 µm
<1 (units/mL)
Colloids–Silica
<10 ppb
Bacteria
<1 (cfu/mL)
* Iron/transition metal content not specified for ASTM Type
I water
Table B-1. ASTM Filtered, Type I Deionized Water Specifications
for Ion Chromatography
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...