3 • Operation and Maintenance
3.1
Getting Started
3.1.1
Initial Startup
1. If you are starting the Thermo Scientific Dionex ICS Series
Photodiode Array Detector (Dionex PDA) for the first time, press the
main power switch on the rear panel (see
2. Press the power button on the front of the detector (see
After initial startup, leave the main power switch on unless instructed
to turn it off (for example, before performing a service procedure).
If the detector is installed with a Dionex ICS-5000 or Dionex ICS-
3000 system, use the power button on the front of the Dionex Dual
Pump (DP) or Dionex Single Pump (SP) for routine on/off control. To
turn off the pump, press and hold the power button for 2 seconds.
If the detector is
not
installed with a Dionex ICS-5000 or Dionex
ICS-3000 system, use the power button on the front of the Dionex
PDA for routine on/off control.
3. Check the status of the LEDs on the front of the detector: The
following LEDs should be lighted:
POWER
,
DEUTERIUM – UV
, and/or
TUNGSTEN – VIS
LED.
4. Wait 20 to 30 minutes for the lamps to stabilize. (For sensitive
applications work or for detector validation, allow at least 2 hours.
After installing a new lamp, allow 8 to 24 hours.)
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...