4 • Troubleshooting
Doc. 065378-01 1/12
63
When the data collection rate is 20 Hz, select parameters that meet the
following criteria:
Where:
O
= Reference wavelength
BW = Reference bandwidth
For example, the following parameters are VALID:
Wavelength = 190 to 500 nm
Reference wavelength = 450 nm
Reference bandwidth = 50 nm
These parameters are valid because 190 to 500 nm is a range of
d
310 nm and
450 nm + 25 nm (50% of 50 nm) = 475 nm, which is less than 500 nm.
The following parameters are NOT VALID:
Wavelength = 190 to 600 nm
Reference wavelength = 750 nm
Reference bandwidth = 50 nm
These parameters are invalid because 190 to 600 nm is a range of more than
310 nm; also, 750 nm + 25 nm (50% of 50 nm) = 775 nm, which is greater
than 600 nm.
To troubleshoot:
Chromeleon cannot perform digital data acquisition of a 3D data field unless
the software includes the 3D Data Acquisition license. Contact Thermo Fisher
Scientific if you want to order this option.
Acquisition On for 3DFIELD rejected—no 3D Data Acquisition
license found.
O
O
O
Maximum
BW
Minimum
d
r
d
2
nm
Minimum
Maximum
310
d
O
O
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...