![Thermo Scientific 074114 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/thermo-scientific/074114/074114_operators-manual_1105368026.webp)
Dionex PDA Operator’s Manual
18
Doc. 065378-01 1/12
Semi-Micro Flow Cell
The semi-micro flow cell assembly (see
) is available in PEEK
and 316 stainless steel. Two built-in heat exchangers—one on the cell
inlet tubing and one on the cell handle—help stabilize the temperature of
the mobile phase before it enters the cell.
The semi-micro flow cell is designed for applications with 2- to 3-mm ID
separator columns and flow rates up to 1.0 mL/min.
Figure 2-6. Semi-Micro Flow Cell Assembly
Semi-Micro Cell Component
PEEK Cell
SST Cell
Inlet tubing: 0.12-mm (0.005-in) ID
P/N 063897
N/A
Outlet tubing: 0.38-mm (0.015-in) ID
P/N 057304
N/A
Inlet tubing: 0.12-mm (0.005-in) ID
N/A
P/N 063903
Outlet tubing: 0.25-mm (0.010-in) ID
N/A
P/N 051650
Reducing union fitting
P/N 055903
P/N 055902
Reducing Union Fitting
Reducing Union Fitting
Cell Outlet Tubing
Cell Inlet Tubing
Heat Exchangers
Flow Cell
Handle
Spring Clip
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...