B • Installation
Doc. 065378-01 1/12
113
B.3 Installing the Flow Cell
1. Grasp the Dionex PDA front cover by the sides and pull straight off to
remove.
2. Remove the flow cell cover from the inside front panel (see
3.
Standard or semi-micro cell
: Remove the tubing bracket from the lamp
cover (see
4. Squeeze the handle of the flow cell and insert it straight into the optical bench.
Release the handle, and then pull it forward slightly. If there is a click, the cell
is properly mounted. If there is no click, rotate the handle slightly just until
the cell clicks into place.
5. Thread the flow cell inlet tubing through the tubing guide on the lamp cover
and reinstall the tubing bracket (see
6. Thread the flow cell outlet tubing through the slot on the side of the flow cell
cover. Replace the cover and reconnect the cell inlet and outlet tubing.
7.
Standard or semi-micro cell:
Connect the cell inlet tubing to the reducing
union fitting (P/N 055903 for a PEEK cell; P/N 055902 for a stainless steel
cell) on the heat exchanger tubing (see
Do not touch the cell windows. If you touch a window, clean it with
isopropyl alcohol (IPA) and a clean lens tissue.
Do not use a PEEK flow cell with normal phase or chlorinated
solvents; these solvents will damage the cell.
Do not use a stainless steel flow cell with low pH solutions; these
solutions may cause corrosion, contamination, and metal leaching.
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...