3 • Operation and Maintenance
Doc. 065378-01 1/12
57
4. After calibrating the lamp, check the intensity again. If the reading is still
too low, follow the steps in
to resolve the problem.
•
Run the wavelength verification test for both the deuterium and tungsten
lamps approximately every 6 months (see
). If a lamp fails the
test, calibrate the wavelength (see
) and then run the
wavelength verification test again.
If a lamp fails the test again, replace it. For lamp replacement instructions, see
for the deuterium lamp or tungsten lamp,
respectively.