Dionex PDA Operator’s Manual
30
Doc. 065378-01 1/12
2.7.2
3D Data Acquisition License
To take full advantage of the Dionex PDA’s capabilities, order
Chromeleon with the 3D Data Acquisition license. This license is
required in order to perform 3D data presentation and spectral analysis
(real-time or post-run), peak purity analysis with selectable criteria, and
spectral library search for positive peak identification.
Chromeleon supports a maximum of two detectors that can record a 3D
field (for example, photodiode array or electrochemical detectors) on one
Chromeleon 7 instrument controller or Chromeleon 6.8 instrument server.
The total number of 3D data channels one instrument controller or server
can acquire simultaneously depends on computer performance.
A 3D Data Acquisition license is not required to display and process data,
only to acquire it. Thus, data acquired on a station equipped with the 3D
Data Acquisition license can be reviewed and processed on any
Chromeleon client station.
2.7.3
System Wellness and Predictive Performance
System Wellness
monitors the overall “health” of the chromatographic
system. It provides built-in diagnostic and calibration features that help
prevent unscheduled system shutdowns and assure reliable operation of
system devices.
Predictive Performance
provides various functions for estimating the
lifetime of replaceable parts and for planning and recording service and
qualification procedures.
A partial list of the System Wellness and Predictive Performance features
is provided below. Refer to the Chromeleon Help for a complete list of the
available features.
•
Recording of detector properties, including the serial number,
firmware version, and operating hours
•
Monitoring of lamp properties, including lamp ignitions, lamp
operation time, and lamp intensity
•
Setting of limits for the lamp intensity and lamp operation time,
including display of warning messages when limits are exceeded
•
Reminders for service and qualification periods, including display of
warning messages when limits are exceeded
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...