9. Squeeze the clip on the UV lamp connector and disconnect it from the 6-pin
bulkhead connector on the side wall (see
10. Loosen the three screws in the deuterium lamp flange (see
) and
pull the lamp out of the optical bench.
NOTE The screws are captive in the lamp base and do not need
to be removed.
11. Inspect the new deuterium lamp (P/N 939016T) for fingerprints and dust. If
necessary, clean the lamp with IPA (isopropyl alcohol).
12. Line up the flange of the new lamp with the two locating pins and three
threaded holes in the optical bench (see
). When the flange is
aligned, gently push the lamp into the optical bench.
13. When the lamp is fully seated, tighten the screws in the flange with a
screwdriver.
14. Squeeze the clip on the UV lamp connector and insert it into the 6-pin
bulkhead connector on the side wall. Route the wires through the wire guide
in the optical bench (see
Figure 5-3. Lamp Connections
To ensure proper performance, the lamp must be fully seated.