Dionex PDA Operator’s Manual
104
Doc. 065378-01 1/12
5. Replace the two fuses in the holder with new 4.0 amp fast-blow IEC 127 fuses
(P/N 954763). Thermo Fisher Scientific recommends always replacing both
fuses.
6. Reinsert the fuse holder into the compartment. (The fuse holder is keyed to fit
only in the correct orientation.) Apply pressure evenly against the holder until
the locks are engaged. When correctly installed, the fuse holder is flush
against the rear panel.
7. Reconnect the power cord and turn on the power.