Dionex PDA Operator’s Manual
68
Doc. 065378-01 1/12
To troubleshoot:
Locate the source of the leak and tighten (or replace) the liquid line
connection. For tightening requirements, refer to
Installation of Dionex
Liquid Line Fittings
(Document No. 031432). The manual is on the Thermo
Scientific Reference Library DVD (P/N 053891).
If the cell is leaking, check that the backpressure on the cell does not exceed
the specification (see
Wipe up all liquid and dry the leak sensor before resuming operation.
If the cell leaks after you resume operation, replace it (see
To troubleshoot:
Follow the troubleshooting steps in
To troubleshoot:
The fan that cools the optical bench may have malfunctioned or a component
on the lamp supply board may have failed. Turn off the Dionex PDA power
and contact Thermo Fisher Scientific for assistance.
NOTE Do not attempt to service the Dionex PDA electronics
components. All repairs of the electronics components
must be performed by Thermo Fisher Scientific
personnel.
Leak sensor error.
Not enough light to perform operation.
Over temperature error.
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...