Dionex PDA Operator’s Manual
70
Doc. 065378-01 1/12
To troubleshoot:
1. Check that the lamp referred to in the error message is turned on. When a
lamp is on, the corresponding LED (
DEUTERIUM – UV
or
TUNGSTEN –
VIS
) is lighted. If necessary, turn on the lamp from Chromeleon or the
TTL input.
2. If you recently installed a new deuterium lamp, check that the UV lamp
connector wires are routed through the wire guide in the optical bench
(see
) and are not pinched under the lamp cover.
3. Check the number of hours the lamp has been in service:
Open the Chromeleon 7 Command window (see
) or the
Chromeleon 6.8 Commands dialog box (see
), and then select
UV
.
If the error message relates to the deuterium lamp, execute the
UVLampAge
command.
UV lamp error.
UV lamp error or control circuitry error while lamp is on.
UV lamp failed to turn on for the required operation.
UV lamp error while attempting to turn on.
Visible lamp error.
Visible lamp error or control circuitry error while lamp is on.
Visible lamp error while attempting to turn on.
Visible lamp failed to turn on for the required operation.
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...