BIOSTRENGTH LINE: Service & Maintenance manual - rev. 2.1
Page 6.29
(1)
Insert the oscilloscope probes between pins 5 (signal) and 6 of the CN21 patch cord. With the
motor enabled you should see a square wave signal ranging from approximately 3.5 Vdc to
11.2 Vdc.
(2)
Insert the oscilloscope probes between pins 5 (signal) and 6 on the CN1 connector on the Bio
servo board. With the motor enabled you should see a square wave signal ranging from
approximately 3.5 Vdc to 11.2 Vdc.
(3)
Insert the oscilloscope probes between pins 9 (signal) and 1 of the CN7 connector on the Bio
servo board. With the motor enabled you should see a square wave signal ranging from
approximately 0 Vdc to 5 Vdc.
On Chest Press and Leg Press, the BS-6/2 cable is divided into 2 parts through the
connector CN36. Carry out the test on this connector and, if necessary, replace only the
defective section of the BS-6/2 cable.
(4)
Insert the oscilloscope probes between pins 9 (signal) and 1 of the K4 connector on the CPU
board. With the motor enabled you should see a square wave signal ranging from
approximately 0 Vdc to 5 Vdc.
Содержание BIOSTRENGTH LINE
Страница 1: ...SERVICE MAINTENANCE MANUAL REV 2 1...
Страница 2: ......
Страница 4: ......
Страница 16: ...BIOSTRENGTH LINE Service Maintenance manual rev 2 1 Page xii Page intentionally left blank...
Страница 88: ...BIOSTRENGTH LINE Service Maintenance manual rev 2 1 Page 2 70 Page intentionally left blank...
Страница 99: ...BIOSTRENGTH LINE Service Maintenance manual rev 2 1 Page 3 11 Figure 3 2 5...
Страница 121: ...BIOSTRENGTH LINE Service Maintenance manual rev 2 1 Page 4 1 4 ACCESSORIES The equipment has no accessories...
Страница 122: ...BIOSTRENGTH LINE Service Maintenance manual rev 2 1 Page 4 2 Page intentionally left blank...
Страница 634: ...BIOSTRENGTH LINE Service Maintenance manual rev 2 1 Page 9 26 Page intentionally left blank...
Страница 645: ......