Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 29
Operation (5-Lead, 5L, LT and XG)
Operation (5-Lead, 5L, LT)
• INTERVAL is set and you press DEFLATE (12):
The timed measurement is suspended and the cuff deflates.
• INTERVAL is set and you change the interval:
The measurement cycle is reset with the new interval. A measurement will be taken after the
new interval has elapsed.
For example, with the interval set to five minutes:
1.3.6
ECG Acquisition
The patient ECG signal is acquired with a patient cable and skin electrodes. The type of skin
electrode and technique of applying the electrodes are major factors in determining the
quality of the signal obtained. Use high quality, silver-silver chloride electrodes. These are
designed to acquire the ECG with excellent base line stability, recovery from defibrillation,
and minimum artifact from patient movement.
WARNING: Ensure that the conductive parts of ECG electrodes do not
contact other conductive parts including earth ground.
1.
For optimal skin contact, thoroughly prep patient skin for electrode placement by doing
the following.
• Shave hair from electrode sites.
• Cleanse skin thoroughly with alcohol to remove skin oils.
• Dry with a rough towel or gauze to remove dry skin.
• Attach electrodes to lead wire first before placing onto patient.
NOTE:
Using more than one type of electrode on the same patient
should be avoided because of variations in electrical
resistance.
TABLE 1-6 Interval Sequence
TIME
MODE/TIMER INTERACTION
RESULT
9:45
Unit turned on with interval set to 5 minutes.
No measurement taken
9:52
Start key pressed
Measurement taken
9:55
Timer requests a measurement
Measurement taken
10:00
Not in Deflate mode and timer requests a measurement
Measurement taken
10:04
Deflate mode is entered
Deflate message is
displayed on display
10:05
Timer requests a measurement
Measurement is skipped
10:07
Interval is changed to 10 min.
Interval timing is reset
10:17
Timer requests a measurement
Measurement taken
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...