Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
7 - 21
Calibration
Initial Set-Up
3.
Verify proper range of the front panel contrast control:
• When the contrast is positioned full up, the LCD display should be completely white.
• When the contrast is positioned full down, the LCD display should be completely blue.
• Set contrast for optimum visibility.
4.
Check the upper Left side of the LCD display for the correct time and date. If incorrect,
follow the steps below top enter the user configuration menu.
• Turn PASSPORT off, and then back on.
• At the moment the LCD displays the message “Selftest Complete”, depress and hold
the “FREEZE” KEY. The LCD displays the USER CONFIGURATION menu.
• Use the SETUP UP ARROW/DOWN ARROW keys to select options, then press select.
Follow the on screen directions to change the date, time and trend setup as required.
Exit the configuration menu by pressing and holding the END key for three seconds.
7.7
Initial Set-Up
1.
Using a patient simulator, connect the ECG, IBP1, IBP2 and temperature cables to the
left side panel. Set the ECG simulator for 60 bpm, 1mv, QRS signal.
2.
Setup the Passport as follows:
• Adult mode (Patient option)
• Lead 1
• ECG size to 1.0cm/mv.
• Speed to 50mm/sec (Setup option)
• Waveform 2 to Cascade. (Setup Option)
7.8
ECG Tests
1.
Observe that the trace display sweeps across the Waveform 1 screen in two seconds.
There should be two complete ECG cycles. The same display and timing should be seen
in the Waveform 2 screen.
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...