Controls, Indicators and Connectors (5-Lead, 5L, LT)
Operation (5-Lead, 5L, LT and XG)
1 - 6
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
12. DEFLATE (NIBP)
A single action key which stops any NIBP measurement in progress, including any timed
measurement sequence, and deflates the cuff. When a timed measurement is stopped an
“NIBP: DEFLATE” message followed by an “NIBP: IDLE” message displays in the NIBP
message window until the timer mode is restarted by either pressing START or changing the
interval.
13. VOLUME (Alarms and Alarm LED)
A repeat action key which adjusts the alarm volume in 5 steps. There is no off position for this
tone. The tone wraps to the minimum volume once the maximum is reached. The red alarm
bell LED illuminates when an alarm condition exists. Refer to “Alarms” on page 1-49.
14. CHECK/RETURN (Alarms)
A single action key which provides access to the alarm menu to view and select alarm
values. Press SELECT to change the alarm values. Pressing CHECK/RETURN a second time
or the END key, returns the display to the main screen and exits the alarm menu.
15. MUTE (Alarms)
A single action and delayed action key which silences all currently alarming parameters or
suspends all alarms. A single key press will silence the current alarm tone for 2 minutes. Any
new alarms that occur while the alarm tone is muted will disable the mute and sound the
alarm tone. An alarm mute symbol is displayed next to each muted parameter. The word
Mute is displayed above the menu selections.
When the key is pressed and held for 3 seconds, all set alarms will be suspended for 2
minutes. This is indicated by the Alarm Mute Symbol and the words All Mute displayed
above the menu selections.
When “Aud alm standby” is set to ON (in the User Configuration Menu), and the MUTE key
is pressed and held for 4 seconds, all alarms are indefinitely suspended. This is indicated by
the Alarm Mute Symbol in all parameter windows and the words Aud Alm Sby flashing
above the menu selections.
The All Mute and Aud Alm Sby modes can be exited by pressing the mute key.
NOTE:
When “Aud alm standby” is set to ON and the unit is power
off and then on again, the unit will power up in the Aud alm
standby mode.
16. TRENDS/RETURN (Trends)
A single action and delayed action key which displays and clears trended data. The most
recent page of data is displayed when the key is first pressed. Press this key a second time to
view additional trended data. Pressing and holding the key for 3 seconds clears all trended
data.
17. TIME UP
p
and
q
TIME DOWN (Trends)
Two repeat action keys which are used to scroll through the trend data.
Alarm Mute
Symbol
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...