Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 133
Operation (5-Lead, 5L, LT and XG)
Operation (XG)
1.6.11.2
CO
2
Waveform
When used with the optional mainstream/sidestream CO
2
or optional Gas Module the
Passport XG may use the breath waveform to report the respiration rate by measuring the
actual breaths per unit time. The advantage of the CO
2
waveform method is that the signal is
a direct result of respiration and can only occur if the patient is actually breathing. Refer to
Section 1.6.12 on page 1-134 through Section 1.6.14 on page 1-146 for more details on
the above options for CO
2
monitoring.