Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 13
Operation (5-Lead, 5L, LT and XG)
Operation (5-Lead, 5L, LT)
1.3
Operation (5-Lead, 5L, LT)
Abbreviated Operating Check List
CAUTION:
Only use the Abbreviated Operating Check List if you are
already familiar with this product. If not, please continue
with the remainder of this chapter, Detailed Operating
Instructions.
A. Setting-Up
1.
Set POWER switch to OFF.
2.
Connect, if desired, peripheral equipment (i.e., Remote Color Display, P.C., etc.).
3.
Attach power pack and/or install charged batteries as needed.
4.
Set POWER switch to ON.
5.
Using the menus and keyboard, set (when appropriate) the following:
--
B. Initiating an NIBP Measurement
1.
Select cuff.
2.
Attach cuff hose to NIBP connector and place cuff on patient.
3.
Select timer interval, if desired.
4.
Select cuff pressure, if necessary.
5.
Press START to begin an NIBP measurement.
6.
Press DEFLATE to suspend measurement.
C. Establishing SpO
2
1.
Select the appropriate sensor.
2.
Attach sensor to Sp
O
2
connector and apply to the patient.
3.
Set either waveform 2 or 3 in the set-up menu to display the Sp
O
2
waveform, if desired.
• Patient Size and Trend Clear
• HR Information
• Set Up Information (speed, waveform
• Sp
O
2
Information 2 and 3, etc.)
• ECG Information
• Recorder Information (when available)
• Resp. Information
• Alarm Limits
• Alarm Volume
• NIBP Information
• Beep Volume
• IBP Information
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...