Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
3 - 5
Specifications
Performance Specifications
Accuracy at 25 °C ambient following
90 second warmup:
±0.1°C 15 to 45 °C - (Plus Probe Error)
±0.2 °F 59 to 113 °F - (Plus Probe Error)
• With Masimo SpO
2
*
SpO2 Accuracy - Saturation
during no motion conditions:
Adult:
70 - 100 % ±2 digits SpO2 (1 S.D.)***
0 - 69 % Unspecified
Neonatal:
70 - 100 % ±3 digits
SpO
2
Accuracy - Saturation
during motion conditions**:
Adult:
70 - 100 % ±3 digits SpO2 (1 S.D.)
0 - 69 % Unspecified
Drift:
None
SpO2 Response Time:
18 seconds to 95% of final step change of
% SpO2 value from 60 - 95% (at 72 bpm)
and with Post Average Time set to 8 seconds
*
This feature applicable only if available / installed in your unit.
** The Masimo option and LNOP-Adt sensors has been validated for motion accuracy in human blood studies
on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping motions at 2
to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to
3 cm in induced hypoxia studies in the range of 70 to 100 % SpO2 against a laboratory co-oximter and ECG
monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation
encompasses 68% of the population.
*** The SPO
2
accuracy during motion conditions is not specificed for the LNOP Ear Sensors. The LNOP Ear Sen-
sors have an SPO
2
accuracy of 70% to 100% ± 3.5% for adults during no motion conditions, however since
the monitor cannot display 1/2 digits, the accuracy shall be rounded to ± 4 digits.
• Respiration (ECG)
Rate Range:
4-200 bpm
Rate Accuracy:
±2 bpm or ±2%, whichever is greater
Minimum Breath
Function of Display Scale. Minimum breath height
Detected:
detected is 0.1 ohm
• Recorder
Speed:
50, 25, 12.5, 6.25, 3.125 mm/sec ± 5%
NOTE:
6.25 and 3.125 are for Respiration only.
Chart Size:
50 mm ± 2 mm
• Battery Performance (Passport 5-Lead, 5L & LT)
Run Time:
2 hours from full charge of new battery
(5-Lead, 5L, & LT)
at 25 °C with SpO2, 1 NIBP every 15
minutes and no recording.
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...