Operation (XG)
Operation (5-Lead, 5L, LT and XG)
1 - 150
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
1.6.15.2
Alarm Limits
All of the alarm limits have an "OFF" position with the exception of low SpO
2
and apnea in
the neonate mode. A separate table of alarm limit settings is maintained for each patient
size. When the patient size is changed, the appropriate table is
automatically used. See table below for alarm ranges.
NOTE:
The alarm parameters that are highlighted in grey are
available only in models that are equipped with the
invasive pressure option.
NOTE:
The ETCO
2
alarm is available only in models that are
equipped with the CO
2
or Gas Module option.
*
These alarm parameters may be set outside the accurate measurement range. Refer to the specifications,
Chapter 3, for accuracy ranges.
®
Respiration rate measurement range is limited to 2 - 150 rpm when Mainstream CO
2
is selected as the rate
source. Values above 150 rpm will be displayed as 150 rpm.
NOTE:
When equipped with the optional Gas Module, a second
and third page of alarm limits are available by pressing the
limit key a second and third time. These Alarm Limits are not
patient size dependent. If the patient size is changed the
Alarm Limits do not change.
ALARM PARAMETERS
HIGH
LO
Parameters
Adult
Ped/Neonate
Adult
Ped/Neonate
Heart Rate (bpm)
Off, 100-250
Off, 100-250
Off, 30-100
Off, 30-100
IBP1 Sys (mmHg)
Off, 70-240
Off, 40-180
Off, 5-130
Off, 5-130
IBP1 Dia(mmHg)
Off, 40-130
Off, 50-100
Off, 5-90
Off, 5-50
NIBP Sys(mmHg)
Off, 70-240
Off, 40-180*
Off, 50-150
Off, 15-130
NIBP Dia(mmHg)
Off, 40-130
Off, 50-100
Off, 30-120
Off, 10-50
IBP2 Mean (mmHg)
Off, 5-150
Off, 5-100
Off, 2-100
Off, 2-50
SpO
2
(%)
Off, 85-100
Off, 85-100
50*-99
50*-99
Resp Rate (rpm)
Off, 30-200*
®
Off, 30-200*
®
Off, 5-50
Off, 5-50
Apnea Delay (sec)
Off, 10-40
Off, 10-30 (Ped) / 10 - 20 (Neo)
ETCO
2
(Torr)
Off, 20-100
Off, 20-100
Off, 5-60
Off, 5-60
ETCO
2
(%)
Off, 2-11
Off, 2-11
Off, 1-8
Off, 1-8
ETCO
2
(kPa)
Off, 2.0-11.0
Off, 2.0-11.0
Off, 1.0-8.0
Off, 1.0-8.0
TABLE 1-1 Alarm Parameters (Gas Module Only)
PARAMETERS
HIGH
LO
Resp Rate*
Off, 20 - 60
Off, 5 - 40
ET CO
2
(kPa or %)
Off, 6.0 - 15.0
Off, 0.5 - 9
ET CO
2
(Torr or mmHg)
Off, 50 - 115
Off, 5 - 70
*
This Alarm Limit is found on page one of the Alarm Limit Menus, since it applies whether the
Gas Module is configured or not. However, when the Gas Module is configured this Alarm
parameter will have different ranges and it will be saved separately.
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...