Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 111
Operation (5-Lead, 5L, LT and XG)
Operation (XG)
A cuff that is too narrow for the limb will result in erroneously high readings. The correct size
of the pressure cuff for a given patient has, among other considerations, a direct bearing on
the accuracy of the obtained NIBP measurements. Base your selection of the cuff size on the
limb circumference of the patient. The table above indicates the available Datascope cuffs for
use with the Datascope Passport XG Monitor. The design dimensions of the cuffs and their
intended uses are based on recommendations of the American Heart Association.
NOTE:
See Accessories section in P/N’s 0070-00-00397, 0070-00-
0440, and 0070-00-0503 for a detailed list of cuffs.
NOTE:
Cuffs become more supple as they age and sometimes
develop permanent folds that can leave temporary marks
on the limb. Any cuffs that exhibit this effect should be
replaced.
NOTE:
Ensure that the pressure tubes are not compressed or
restricted.
*
When using the thigh cuff, this product will not comply with AAMI accuracy standards.
**
The limb circumferences of the Color Coded Cuffs adhere to the AHA guidelines for size.
The pressure on the limb may not fall to zero between measurements if the cuff is wrapped
too tightly. Therefore, insure that the cuff is properly applied.
The skin is sometimes fragile (i.e., on pediatrics, geriatrics, etc.). In these cases, a longer
timer interval should be considered to decrease the number of cuff inflations over a period of
time. In extreme cases, a thin layer of soft roll or webril cotton padding may be applied to
the limb in order to cushion the skin when the cuff is inflated. This measure may affect NIBP
performance and should be used with caution.
2.
Attach cuff hose to CUFF Connector (27).
3.
Apply the cuff to the patient. To reduce errors, the cuff should be fitted snugly, with little
or no air present within the cuff. Be sure the cuff lies directly against the patient’s skin.
No clothing should come between the patient and the cuff.
NOTE:
The NIBP cuff should not be placed on a limb that is being
utilized for any other medical procedure. For example, an
I.V. catheter or an SpO
2
sensor.
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...