Operation (XG)
Operation (5-Lead, 5L, LT and XG)
1 - 166
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
GM: AGENT
MISMATCH - ENF
Appears in Inverse Video when the system
detects Enflurane as the primary agent and the
manually selected agent is neither Enflurane
nor the detected secondary agent.
Match the Agent
administered with the
Agent selected, or select
Agent Auto ID.
GM: AGENT
MISMATCH - SEV
Appears in Inverse Video when the system
detects Sevoflurane as the primary agent and
the manually selected agent is neither
Sevoflurane nor the detected secondary agent.
Match the Agent
administered with the
Agent selected, or select
Agent Auto ID.
GM: AGENT
MISMATCH - DES
Appears in Inverse Video when the system
detects Desflurane as the primary agent and
the manually selected agent is neither
Desflurane nor the detected secondary agent.
Match the Agent
administered with the
Agent selected, or select
Agent Auto ID.
GM: UNKNOWN
AGENT
Appears in Inverse Video when the system
detects a gas that does not match the
spectroscopic signatures of the five known
anesthetic agents.
Rectify situation.
GM: CANNOT
ZERO ... RETRYING
Appears when the Passport requests a Zeroing
(either on the automatic cycle or by a user
request), and the Gas Module is unable to
initialize the cycle.
Allow system to retry
without intervention. If
problem persists call for
service.
GM: RESP HIGH
Appears when the respiration rate is greater
than or equal to the value set for the Resp High
Alarm.
Check patient.
GM: RESP LOW
Appears when the respiration rate is less than
or equal to the value set for the Resp Low
Alarm.
Check patient.
GM: ET CO
2
HIGH
Appears when the End Tidal CO
2
measurement is greater than or equal to the
value set for the ET CO
2
High Alarm.
Check patient.
GM: ET CO
2
LOW
Appears when the End Tidal CO
2
measurement is less than or equal to the value
set for the ET CO
2
Low Alarm.
Check patient.
GM: INSP CO
2
HIGH
Appears when the FiCO
2
measurement is
greater than or equal to the value set for the
INSP CO
2
High Alarm.
Check patient.
GM: ET O
2
HIGH
Appears when the End Tidal O
2
measurement
is greater than or equal to the value set for the
ET O
2
High Alarm.
Check patient.
GM: ET O
2
LOW
Appears when the End Tidal O
2
measurement
is less than or equal to the value set for the ET
O
2
Low Alarm.
Check patient.
GM: INSP O
2
HIGH
Appears when the FiO
2
measurement is
greater than or equal to the value set for the
INSP O
2
High Alarm.
Check patient.
GM: INSP O
2
LOW
Appears when the FiO
2
measurement is less
than or equal to the value set for the INSP O
2
Low Alarm.
Check patient.
GM: INSP N
2
O
HIGH
Appears when the Inspired N
2
O measurement
is greater than or equal to the value set for the
INSP N
2
O High Alarm.
Check patient.
GM: ET AGENT
HIGH
Appears when the End Tidal Anesthetic Agent
measurement is greater than or equal to the
value set for the ET Agent High Alarm.
Check patient.
MESSAGE
REASON
SOLUTION
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...