RECORDER DOOR OPEN
The door of the recorder is not closed.
Close recorder door.
RECORDER PAPER OUT
The roll of recorder paper is used up.
Replace with a new roll
of paper.
MESSAGE
REASON
SOLUTION
CO
2
: INITIALIZING
This message usually appears at power
up. The CO
2
subsystem has been reset
and is performing internal self tests prior
to operation.
Wait for the message to
go away. It takes about
15 seconds for the CO
2
subsystem to initialize.
CO
2
: NO SENSOR
No CO
2
sensor is connected to the
monitor.
Attach the CO
2
sensor
to the CO
2
connector
(30).
CO
2
: SENSOR WARMING
UP
The CO
2
sensor has not reached its
operating temperature. Either the
monitor was just turned on or the sensor
was recently plugged in or the sensor
was removed from the adapter or
calibration cells.
Wait for the message to
go away. It takes up to
five minutes for the
sensor to warm up.
CO
2
: CHECK ADAPTER
The CO
2
sensor is not correctly placed
on the airway adapter or a different
type of airway adapter has been
connected or the adapter is dirty or
damaged.
Check the installation
and condition of the
airway adapter. Clean
or replace the adapter if
necessary. An adapter
calibration should be
performed if the
message persists. This is
accessed through the
CO
2
menu.
CO
2
: CHECK SENSOR
The CO
2
sensor is out of calibration or
damaged.
A zero calibration
should be performed if
the message persists.
This is accessed through
the CO
2
menu.
CO
2
: SENSOR ON
REFERENCE CELL
The CO
2
sensor is positioned on the
reference cell. This is located on the
sensor cable adjacent to the connector.
The ETCO
2
reading should range from
36 to 40 Torr when on this cell.
The sensor should be
removed from the
reference cell and
placed back on the
airway adapter.
CO
2
: SENSOR ON ZERO
CELL
The CO
2
sensor is positioned on the
zero cell. This is located on the sensor
cable adjacent to the connector. The
ETCO
2
reading should be 0 when on
this cell.
After completing the
calibration, the sensor
should be removed from
the zero cell and placed
back on the airway
adapter.
CO
2
: ZERO CAL IN
PROGRESS
The CO
2
sensor is placed on the zero
cell and a CO
2
calibration is in
progress.
Wait for the zero
calibration to complete.
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...