1.3.19.17
User Configuration - Audio Alarm Standby Menu
Aud alm standby - When this function is set to ON, pressing and holding the MUTE key for 4
seconds will indefinitely suspend the alarms.
FIGURE 1-46
Audio Alarm Standby Menu
1.3.19.18
User Configuration - Serial Output Type Menu
Serial output type - Select the communication protocols for interfacing with other specialized
equipment.
FIGURE 1-47
Serial Output Type Menu
CHANGE ALARM STANDBY
Aud alm standby:
= Adjust valve
END = quit
SELECT = Enter/move
OFF
Alarm Audio delay
Serial output type
Date
USER CONFIGURATION
Press & hold the END key for 3 sec, to leave configuration mode.
Time
Trend
Choices: ON/OFF
Temp
NIBP: Idle
SPO2
Hr SpO2 size
Aud alm standby
CO2
CHANGE SERIAL OUTPUT TYPE
Serial output type:
= Adjust valve
END = quit
SELECT = Enter/move
Visa
Alarm Audio delay
Date
USER CONFIGURATION
Press & hold the END key for 3 sec, to leave configuration mode.
Time
Trend
Choices: Visa,Accutorr,Msg,Diap
Temp
NIBP: Idle
SPO2
Hr SpO2 size
Aud alm standby
Serial output type
CO2
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...