Operation (5-Lead, 5L, LT)
Operation (5-Lead, 5L, LT and XG)
1 - 46
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
5.
Check that the connections have been made correctly by verifying a proper CO
2
waveform (capnogram) on the monitor display.
NOTE:
The CO
2
waveform is displayed by choosing CO
2
as
Waveform 2 or 3 and the Resp. Source in the Setup menu
6.
The sensor cable should face away from the patient. To secure the sensor cable safely
out of the way, attach the sensor cable holding clips to the airway tubing, then connect
the sensor cable to the clips.