Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 27
Operation (5-Lead, 5L, LT and XG)
Operation (5-Lead, 5L, LT)
If NIBP is a selected trend trigger, then NIBP and heart rate values are stored in trend lists.
The NIBP and NIBP heart rate will be automatically removed from the display after 15
minutes have elapsed.
If another heart rate source is selected (ECG, IBP or SpO
2
)< before the 15 minute display
time has elapsed, the NIBP heart rate will be replaced by the heart rate from the selected
source.
To display heart rate from NIBP:
a.
Choose AUTO for HR source.
b.
Discontinue monitoring of ECG, SpO
2
and IBP.
c.
Set high and low heart rate alarms to off.
d.
Make a valid NIBP measurement.
During or after an NIBP measurement, one of several advisory messages may be
displayed. Refer to “Status Messages” on page 1-57, for their explanations.
7.
If desired, press DEFLATE (12) to interrupt a measurement. The cuff will deflate.
1.3.5.2
Automatic Initiation of NIBP Measurements
Follow steps 1 - 5 in Section 1.3.5.1 on page 1-25.
6.
Press INTERVAL (11) until the desired time displays. The choices are: OFF, continuous,
1, 2.5, 5, 10, 15, 20, 30, 60, and 120 minutes.
7.
A measurement will be taken when the selected interval has elapsed. If an immediate
measurement is desired, press START (10).
NOTE:
If the monitor is in the interval mode when it is turned on,
no measurement will be taken until the “START” key is
depressed. Subsequent measurements are referenced to the
time the unit was turned on. See example in Section 1.3.5.5
on page 1-28.
NOTE:
When the NIBP continuous interval is chosen, the Passport
will continually take back to back blood pressure readings.
As a safety precaution, a five minute limit is placed on
continuous measurements. After 5 minutes, the NIBP
interval will automatically switch to measurements taken
every 5 minutes.
Automatic Adjustment in the Timer Mode
In the timer mode, the unit adjusts the inflation pressure according to the previous reading of
the systolic pressure. After the first measurement in the timer mode, the inflation pressure is
the previous sy50 mmHg in the Adult Mode and +30 mmHg in the pediatric and
neonate mode.
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...