Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
3 - 7
Specifications
Performance Specifications
• CO2 - Mainstream and Sidestream (0998-00-0134-XX, 0998-00-0137-XX
units only)
Range:
0-100 TORR, CO2 partial pressure
Mainstream / Sidestream
(Non-Intubated) Accuracy:
±2 TORR (0-40 TORR)
±5% (41-70 TORR)
±8% (71-100 TORR)
Sidestream
(Intubated) Accuracy*:
±5 Torr (0 - 40 Torr)
±13% (41 - 100 Torr)
Respiration Rate Range:
1-150 bpm (mainstream)
1-50 bpm (sidestream**)
Respiration Rate Accuracy:
±1 bpm
CO2 waveform recognition
15 seconds for waveform
Time to meet full specs:
1 minute to meet spec
Sensor warmup time:
Less than 5 minutes
Barometric Pressure Error:
±2 Torr max additional, 564 to 760 mmHg
*
Test conditions as follows:
8 to 18 bpm with 1000cc tidal volume with max PEEP of 15cm-H2O
18 to 30 bpm with 400cc tidal volume with max PEEP of 15cm-H2O
30 to 50 bpm with 70cc tidal volume with max PEEP of 5cm-H2O
Sidestream Flow Rate:
180 ±20 cc/minute:
Reference Cell Reading:
36 to 40 TORR
Sensor on Zero Cell:
0 ±1 TORR
Zero Calibration:
Less than 30 seconds
Adapter Calibration:
Less than 30 seconds
Pump Calibration:
Less than 30 seconds
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...