Sidestream Pump Calibration
Calibration
7 - 28
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
FIGURE 7-14
Sidestream CO
2
Control Module
7.14.2
Sidestream Adapter Calibration
Performing a Pump Calibration automatically calibrates the adapter. An Adapter Calibration
is necessary whenever the sampling cell airway adapter is replaced or when the “Check
Adapter” message displays.
To perform an adapter calibration:
1.
Plug the Capnostat® CO
2
sensor into the Passport XG CO
2
connector (34). The “CO2:
Sensor Warming Up” message is displayed. The adapter calibration cannot be
performed until the “CO2: Sensor Warming Up” message disappears.
2.
Ensure the nasal cannula is away from all sources of CO
2
(including the patient’s and
your own exhaled breath, and ventilator exhaust valves).
3.
Choose Start Adapter Cal - Yes from the CO
2
menu. (See “Use of Menus” on
page 109.)
NOTE:
If the monitor detects changing CO2 levels (breaths) during
a adapter calibration, a “Breaths Detected ... Retry”
message displays, and then an “Adapter Cal Failure”
message will display. Remove the source of CO2 and repeat
the calibration.
VR2
IC3
J201
J103
IC6
J202
IC7
IC9
IC2
IC13
IC1
IC12
IC4
IC5
IC11
IC10
VR1
Adjust VR2 to 180 cc/min +/- 20 cc/min
Sidestream CO2 Control Module
Y1
Y2
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...