Operation (XG)
Operation (5-Lead, 5L, LT and XG)
1 - 98
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
D.
Initiating an IBP Measurement
1.
Open the transducer stop-cock to atmosphere.
2.
Press and hold ZERO1 or ZERO2 for 1 second until “beep” sounds.
3.
Turn stop-cock to patient to begin monitoring.
E. Establishing SpO
2
1.
Select the appropriate sensor.
2.
Attach sensor to SpO
2
connector and apply to the patient.
3.
Set either waveform 2 or 3 in the set-up menu to display the SpO
2
waveform, if
desired.
F. Establishing CO
2
/ CAPNOSTAT®*
1.
Plug sensor into the side of the monitor.
2.
If a new CAPNOSTAT is used, calibrate the system.
3.
Apply sensor to the breathing circuit.
4.
Set Waveform 2 or 3 to CO
2
or Auto Display CO
2
to Waveform 2 or 3.
5.
Set the Resp. Source to CO
2
/ECG.
*
CAPNOSTAT® is a trademark of NOVAMETRIX, Inc.
G. Recording Information
1.
Select wave to be recorded through Record Menu. Press PRINT to start Recording.
2.
Press PRINT again to stop the recording in process.
3.
If a printout of a tabular trend is desired, press DISPLAY followed by PRINT.
H. Temperature Measurement
1.
Attach the desired temperature probe to the Temperature connector on the side of the
Passport XG. The Passport XG automatically detects which probe is connected, 400
or 700 series.
I.
CO
2
, O
2
, N
2
O and Agent Monitoring with the Gas Module
1.
Connect nasal cannula (non-intubated) or sample line (intubated) to the Patient.
2.
Connect other end of nasal cannula or sample line to Gas Module.
3.
Empty and clean water trap if more than half full or whenever changing Patients.
4.
Set Waveform 2 or 3 to CO
2
, Agent or O
2
or Auto Display CO
2
to Waveform 2 or
3.
5.
Set the Resp. Source to CO
2
/ECG.
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...