Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 155
Operation (5-Lead, 5L, LT and XG)
Operation (XG)
The Temperature scale can be changed between Fahrenheit and Centigrade.
The HR SpO
2
size change option allows the user to select a large SpO
2
reading compared
to HR or vice versa.
The Alarm Audio Standby allows the user to enable (or disable) this feature (see “* This
Alarm Limit is found on page one of the Alarm Limit Menus, since it applies whether the Gas
Module is configured or not. However, when the Gas Module is configured this Alarm
parameter will have different ranges and it will be saved separately.” on page 1-150).
The Serial Output Type allows the user to select the communication protocols for interfacing
with other specialized equipment: VISA, VISA with Admit from bedside feature, Accutorr
(uses the same data protocol as an Accutorr), message (for diagnostic purposes), Gas
Module (optional) with the Passport XG Monitor. The baud rate for the Accutorr protocol can
be set for 1200, 2400, 4800 or 9600 bps.
Once the unit is in the User Configuration Mode, the time and/or date can be changed using
the set-up keys as described in “Use of Menus” on page 1-109.
1.6.18
Recorder (optional)
The Datascope Passport XG Recorder can provide a permanent record of a patient’s: systolic
pressure, diastolic pressure, mean pressure, heart rate, SpO
2
,CO
2
, O
2
, Agent, respiration
and temperature. It is a two trace thermal strip chart recorder with an integral paper spool.
The recorder uses plain white thermal paper 5 cm wide (see “Recorder Paper Replacement”
on page 8-5 for replacement instructions).
All grid patterns and data are printed by the recorder.
1.6.18.1
Operation of Printer
1.
Using the UP
q
and
q
DOWN arrow keys (1), select RECORDER from the main
menu.When RECORDER is highlighted a RECORDER MENU is displayed in the multi-
function area. A sample of this menu is shown to the right.
2.
Press SELECT (2) to enter the RECORDER MENU.
3.
Using the UP
q
and
q
DOWN arrow keys (1) and SELECT (2) keys, (see “Use of
Menus” on page 1-109) set the desired waveforms or trend to be printed.
4.
Press PRINT (22) to initiate a printing or stop a printing when one is in progress.
When PRINT (22) is pressed to initiate a printing, a 16 second strip is printed. The 16
second strip consists of 8 seconds of prior and 8 seconds of post waveform from when PRINT
(22) is pressed. If a continuous printing is required, press and hold PRINT (22) for 3 seconds
(until a beep is heard). Press PRINT again to stop a real time printing.
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...