Operation (5-Lead, 5L, LT and XG)
Operation (XG)
7.
If desired, press DEFLATE (15) to interrupt a measurement. The cuff will deflate.
1.6.5.2
Automatic Initiation of NIBP Measurements
Follow steps 1 - 5 in the “Automatic Initiation of NIBP Measurements” on page 1-113.
6.
Press INTERVAL (14) until the desired time displays. The choices are: OFF, continuous, 1,
2.5, 5, 10, 15, 20, 30, 60, and 120 minutes.
7.
A measurement will be taken when the selected interval has elapsed. If an
immediate measurement is desired, press START (13).
NOTE:
If the monitor is in the interval mode when it is turned on,
no measurement will be taken until the “START” key is
pressed. Subsequent measurements are referenced to the
time the unit was turned on. See example in “Start and
Deflate Functions” on page 1-114.
NOTE:
When the NIBP continuous interval is chosen, the Passport
XG will continually take back to back blood pressure
readings. As a safety precaution, a five minute limit is
placed on continuous measurements. After 5 minutes, the
NIBP interval will automatically switch to measurements
taken every 5 minutes.
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...