Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
3 - 1
3.0
Specifications
Contents of this Chapter ............................................................. Page
3.1 Performance Specifications ................................................. 3-1
3.2 Power Pack Electrical Ratings ............................................. 3-8
3.3 Environmental Conditions ................................................... 3-8
3.4 Physical Characteristics ....................................................... 3-8
3.5 Agency Compliance ............................................................ 3-8
3.1
Performance Specifications
• ECG
Leads:
I, II, III, aVR, aVL, aVF, V
Display Sensitivity:
0.25, 0.5, 1.0, 2.0, 3.0, 4.0 cm/mV ±10%
Frequency Response to Screen:
ESU Filter Out:
0.5 to 40 Hz, -3db max @ 1 mVpp input
(notch filter disabled).
ESU filter in:
0.5 to 20 Hz -3db max @ 1 mVpp input
CMRR:
90 dB min. maximum output of 1 mV p-p
(RTI) over 60 second period with a 51 K
ohm lead imbalance and ± 300 mV dc
offset. ANSI/AAMI EC13-1992.
Defibrillator Overload
Protection:
Withstand Lown and trapezoidal or Edmark
waveform up to 400 joules applied to 100
ohm load and electrode at 20 second
intervals. As per ANSI/AAMI EC13-1983
par. 3.2.2.2.
Recovery Time:
Time for trace to return to screen from
defibrillator mono pulse 3 sec. maximum
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...