Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 59
Operation (5-Lead, 5L, LT and XG)
Operation (5-Lead, 5L, LT)
• Arm should be at heart level.
*
The presence of arrhythmias may increase the time required to complete a measurement and may extend
this time to a point where a measurement cannot be completed.
1.3.16.2
SpO
2
Messages
The following messages pertain to Nellcor SpO
2
Operation.
MESSAGE
REASON
SOLUTION
SpO
2
NO SENSOR
No sensor is connected to the monitor.
Attach sensor cable to
the monitor.
SpO
2
SENSOR OFF
Sensor may not be connected to the
patient.
Check patient
connection.
SpO
2
INTERFERENCE
Noise detected on the pulse signal
prevents pulse discrimination.
Decrease patient
motion, check sensor.
SpO
2
PULSE SEARCH
Hardware settings are being adjusted in
order to discriminate a pulse waveform.
Change to site where
pulse is stronger if
patient is
vasoconstricted.
Change or readjust
sensor if loose.
SpO
2
WEAK PULSE
Detected patient pulse is marginal.
Change site.
SpO
2
NO PULSE
No detectable pulse is measured.
CHECK SpO
2
SENSOR
Insufficient light is received by the sensor
detector, or the sensor has a malfunction.
Check alignment of
emitter and detector or
change sensor.
SpO
2
PR UNDER 30
Detected pulse rate is below thirty beats
per minute.
Check patient status.
Check alignment of
sensor. Reapply sensor.
SpO
2
PR OVER 250
Detected pulse rate is above 250 beats
per minute.
Check patient status.
Decrease patient
motion. Reapply sensor.
SpO
2
UNCALIBRATED
Detected SpO
2
falls below calibrated
range.
Check patient status.
Check sensor.
SpO
2
SYSTEM CHECK
Self test is being performed.
Wait for completion of
self test.
SpO
2
FAILURE
ROM checksum fail. RAM test fail. Filter
mismatch. Offset mismatch.
Power cycle unit. If
message reappears, call
Service.
MESSAGE
REASON
SOLUTION
SpO
2
SYSTEM
CHECK
Self test is being performed.
Wait for completion of
self test.
SpO
2
NO
SENOSR
No sensor is connected to the monitor.
Attach sensor cable to
the monitor.
SpO
2
PULSE
SEARCH
Hardware settings are being adjusted in order to
discriminate a pulse waveform.
Change to site where
pulse is stronger if
patient is
vasoconstricted.
Change or readjust
sensor if loose.
SpO
2
NO
PULSE
No detectable pulse is measured.
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...