NIBP Measurement Messages
• Always have an alternate method of BP verification available.
• On vasoconstricted patients, failure to evacuate air from the cuff can distort BP
MESSAGE
REASON
SOLUTION
NIBP: IDLE
Displayed while system is idle.
Press START to take a
single measurement.
Press INTERVAL to select
an interval and start
timed measurements.
NIBP: DEFLATE
Displayed while the timer mode is stopped, usually
by pressing the DEFLATE key.
Press START to take an
immediate measurement
and resume timed
measurements.
NIBP: INTERVAL
Displayed in the interval between two timed
measurements.
Press DEFLATE to
suspend timed
measurements. Press
INTERVAL to change
timer to OFF to stop
timer.
NIBP:
INITIALIZING
The NIBP system is being initialized.
Wait until initialization
is complete, indicated
by NIBP: IDLE message.
Initialization may take
up to 3 minutes from
power up.
NIBP: FAILURE
The NIBP system has detected a potential error.
Power cycle unit. If
message reappears, call
Service.
NIBP:MEASURIN
G CUFF
Displayed during a measurement to show the cuff
pressure.
Press DEFLATE to
suspend a measurement
and deflate the cuff.
NIBP:RETRY
MOTION
Measurement has been attempted but no reading
was possible due to detected motion and the retry
limit has not been reached.
Retry will be attempted.
Have patient remain
still.
NIBP:RETRY
PUMP HIGHER
A measurement has been attempted but no reading
was possible due to pulse detection at highest cuff
pressure and the retry limit has not been reached.
Retry will be attempted.
Check that appropriate
patient size is set. Preset
initial inflation pressure.
NIBP:RETRY
A measurement has been attempted but no reading
was possible.
Retry will be attempted.
Check for leaks and
quality of peripheral
pulses. Decrease patient
movement. Switch cuff
to another limb.
NIBP:UNABLE TO
MEASURE*
An unsuccessful measurement cycle has been
completed.
Switch cuff to another
limb. Decrease patient
movement. Press START
to retry. Be prepared to
auscultate BP manually.
NIBP:CUFF
OVERPRESSURE
The internal hardware cuff pressure check valve
has been tripped.
Power cycle unit. If
message reappears, call
Service.
*
The presence of arrhythmias may increase the time required to complete a measurement and may extend
this time to a point where a measurement cannot be completed.
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...