Operation (XG)
Operation (5-Lead, 5L, LT and XG)
1 - 142
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
NOTE:
The CO
2
waveform is displayed by choosing Auto Display
CO
2
as Waveform 2 or 3 in the setup menu and the Resp.
Source as CO
2
/Off or CO
2
/ECG in the Resp menu.
2.
Select CO
2
/Off or CO
2
/ECG as the Resp. Source in the Resp Menu.
3.
Snap a clean airway adapter with tubing into the U-shaped Capnostat® CO
2
sensor.
Align the arrow on the bottom of the adapter with the arrow on the bottom of the sensor.
See Figure 1-30.
NOTE:
Position the airway adapter with it’s windows in a vertical
and NOT a horizontal position. This helps keep patient
secretions from “pooling” on the windows.
4.
Connect the sampling tubing to the CO
2
Pump Input connector (33).
NOTE:
Ensure all tubing connections are secure. When required,
connect a waste gas scavenging system to the Pump
Exhaust fitting (39) via a 1/8" barbed insert (P/N 0103-00-
0454-02).
CAUTION:
Vacuum (negative pressure) should not exceed 1 mmHg at
the Passport Pump Exhaust fitting (39). Excessive scavenge
vacuum may result in an “OCCLUSION” message or
damage to the Passport’s internal pump. The scavenge
system if used must be on during calibration.
5.
Connect the disc filter to the airway adapter.
NOTE:
The disc filter is not recommended for breath rates over 18
bpm, due to the increased dead space.
FIGURE 1-77
Adult/Pediatric Airway Adapter (Intubated)
6.
Connect the five foot capillary tubing to the filter.
7.
Connect the Nafion tube to the capillary tube. Nafion is recommended for high humidity
situations.
8.
Press the CO
2
PUMP key (12), then the SELECT key (2), to turn the sampling pump on
and to initiate a pump calibration. When the pump calibration is complete, the message
“CO
2
: Pump Cal Complete” displays. See “Sidestream CO
2
Messages (on Passports
with Sidestream CO
2
only)” on page 1-164 for additional CO
2
messages.
NOTE:
This pump calibration also calibrates the adapter. It is not
necessary to perform a separate adapter calibration at this
time.
To Monitor
Sampling Inlet
Capnostat CO
2
Sensor (Kit)
P/N 0600-00-0037
Disc Filter
P/N 0103-00-0445-01
Capilary Line, 5 foot
P/N 0683-00-0405-11
Nafion Tubing Assembly
P/N 0008-00-0259
Straight Intubated Adapter
P/N 0683-00-0242-21
To Patient
To Ventillator
Sampling Airway Adapter
with Tubing
P/N 0103-00-0444-01
REF
To Monitor
CO
2
Connector
Capnostat
Serial Number
XXXXX
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...