Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 179
Operation (5-Lead, 5L, LT and XG)
Operation (XG)
1.6.23.5
IBP menu
IBP1 Scale, IBP2 Scale:
Sets the full scale value of the display.
FIGURE 1-97
IBP Menu
1.6.23.6
Resp Menu
Resp Speed:
Sets the speed of the respiration waveform. This may be set either in this
menu or in the Gas Module Menu (if installed).
Scale:
Sets the scale of the Resp waveform.
Resp Source:
Sets the source of the respiration information. “CO
2
/OFF” sets the source to
CO
2
, if CO
2
is present. If CO
2
is not present, the Resp Source is OFF. “CO
2
/ECG” sets the
source to CO
2
, if CO
2
is present. If CO
2
is not present, the Resp Source is set to ECG. If
CO
2
option is not installed, Resp Source choices are OFF or ECG.
CHANGE IBP
IBP1 Scale:
IBP2 Scale:
Choices: 37.5, 75, 150, 300 mmHg
= Adjust valve
EXIT = quit
SELECT = Enter/move
37.5
150
mmHg
CHANGE IBP
IBP1 Scale:
IBP2 Scale:
Choices: 37.5, 75, 150, 300 mmHg
= Adjust valve
EXIT = quit
SELECT = Enter/move
150 mmHg
37.5
mmHg
mmHg
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...