Operation (XG)
Operation (5-Lead, 5L, LT and XG)
1 - 108
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
NOTE:
The sample display shown may not be an exact
representation of your unit.
1.6.3.4
Parameters Display Area
FIGURE 1-64
Parameters Display Area
The parameters display area contains the current values of the patient parameters. The
contents and the layout of the parameters area depends on how many invasive blood
pressure transducers are connected and whether or not the Gas Module option is installed to
the monitor. The example above is with two transducers connected and with the Gas Module
configured.
Within the NIBP parameter area the interval chosen is displayed and the elapsed time (ET)
since the last NIBP measurement was taken is displayed. If another NIBP measurement is not
taken within 15 minutes the ET and the NIBP readings will change to dashes.
An “Artifact” message is displayed in this area, under the ECG Heart Rate, whenever 60 Hz
noise is present on the ECG waveform.
NOTE:
The sample display shown may not be an exact
representation of your unit.
1.6.3.5
Message Display Area
This area of the display contains messages relating to NIBP, SpO
2
, CO
2
, Gas Module and
recorder operation. See “Status Messages” on page 1-158 for a complete list of these
messages.
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...