Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
7 - 23
Calibration
IBP1 and IBP2 Verification
Press the TREND/RETURN key and examine the trend data. The high heart rate value should
be in reverse video indicating the high HR alarm was violated.
7.9
IBP1 and IBP2 Verification
1.
Set the simulator to 0 mmHg for both IBP1 and IBP2.
Set the Passport to display IBP 1 in WINDOW 2 and IBP 2 in WINDOW 3 using the
SETUP menu. Set the pressure scale to 300 mmHg for both IBP 1 and 2.
2.
Zero IBP 1 by pressing the ZERO 1 key. Verify that the systolic, diastolic, and mean
display changes from “dashed lines” to zero +/- 1 mmHg within 2 seconds and the
“TRANSDUCER NOT ZEROED” message is removed. Zero IBP 2 and verify proper
operation.
3.
Apply 50, 150 and 300 mmHg and verify that the following parameters SYS/MEAN/
DIA agree.
4.
Apply a 120/80 mmHg signal to IBP 1 and 60/20 mmHg to IBP 2.
Verify that the reading in the IBP 1 window is 120/80 mmHg +/- 3 mmHg and 60/20
mmHg +/- 2 mmHg in the IBP 2 window. Verify the correct waveforms are displayed on
the LCD display.
7.10
Temperature Verification
1.
Set the simulator to 37º C, 700 series probe.
2.
Verify that the temperature displays is 37.0º +/-0.3.
3.
Set the simulator to 37º C, 400 series probe.
4.
Verify that the temperature displays is 37.0º +/-0.3.
NOTE:
If temperature exceeds 45
º
C or is below 0
º
C, the
temperature display goes to “dashed lines”.
Applied Pressure
Readout
50 mmHg
48 - 52
150 mmHg
147 - 153
300 mmHg
294 - 306
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...