Operation (5-Lead, 5L, LT and XG)
Operation (XG)
NOTES:
For optimal results, DO NOT place the airway adapter between the endotracheal
tube and the elbow, as this may allow patient secretions to block the adapter windows.
Position the airway adapter with it’s windows in a vertical and NOT a horizontal position.
This helps keep patient secretions from “pooling” on the windows.
5.
Check that the connections have been made correctly by verifying a proper CO
2
waveform (capnogram) on the monitor display.
NOTE:
The CO
2
waveform is displayed by choosing Auto Display
CO
2
as Waveform 2 or 3 and the Resp. Source as CO
2
/Off
or CO
2
/ECG in the Resp menu.
6.
The sensor cable should face away from the patient. To secure the sensor cable safely
out of the way, attach the sensor cable holding clips to the airway tubing, then connect
the sensor cable to the clips.
1.6.12.5
Use of Neonatal Airway Adapter
The neonatal airway adapter should be used when monitoring patients with endotracheal
tube diameters less than or equal to 4.0 mm.
1.
Verify the windows are clean and dry. Clean or replace the adapter if necessary.
2.
Snap the airway adapter into the Capnostat® sensor. Align the arrow on the bottom of
the airway adapter with the arrow on the bottom of the Capnostat®. Press the sensor
and airway adapter together until they “click”.
3.
If necessary, perform an adapter calibration. See “Adapter Calibration” on page 1-
135.
4.
Place the Capnostat® / airway adapter assembly between the elbow and the ventilator
circuit wye.
FIGURE 1-73
Neonatal Airway Adapter
NOTES:
For optimal results, DO NOT place the airway adapter between the endotracheal
tube and the elbow, as this may allow patient secretions to block the adapter windows.
Position the airway adapter with it’s windows in a vertical and NOT a horizontal position.
This helps keep patient secretions from “pooling” on the windows.
Neonatal Airway Adapter
CAPNOSTAT
CO Sensor
2
R
To
Patient
To
Ventilator
Wye
Содержание Passport 5-Lead
Страница 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Страница 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Страница 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Страница 544: ...0070 00 0420 Rev T April 13 2011...