background image

3 6

Ultrasonic baths for aqueous 

fl

 uids 

User instructions

Valid for: 

RK 31, RK 31 H, RK 52, RK 52 H
RK 100, RK 100 H, RK 102 H, RK 103 H, RK 106
RK 156, RK 156 BH
RK 170 H
RK 255, RK 255 H
RK 510, RK 510 H, RK 512 H
RK 514, RK 514 H, RK 514 BH
RK 1028, RK 1028 H, RK 1028 C, RK 1028 CH
RK 1040, RK 1050, RK 1050 CH

1556g GB/2019-02

Содержание Sonorex Super RK 31

Страница 1: ...s Valid for RK 31 RK 31 H RK 52 RK 52 H RK 100 RK 100 H RK 102 H RK 103 H RK 106 RK 156 RK 156 BH RK 170 H RK 255 RK 255 H RK 510 RK 510 H RK 512 H RK 514 RK 514 H RK 514 BH RK 1028 RK 1028 H RK 1028...

Страница 2: ...the translation and the original of this document the original version has priority BANDELIN accepts no responsibility or liability for damages caused by improper handling or usage contrary to the in...

Страница 3: ...ns modifications both the warranty claim and the CE conformity will expire If service is required please contact the specialist dealer in charge or the manufacturer Symbols used Symbol Significance Ex...

Страница 4: ...10 2 Preparation 11 2 1 Scope of delivery 11 2 2 Set up assembly 11 2 3 Start up 11 3 Operation 12 3 1 Operating elements 12 3 1 1 Ultrasound 12 3 1 2 Heating 13 4 Use 14 4 1 Instructions for use 14 4...

Страница 5: ...ndelin com 6 Maintenance and repair 22 6 1 Maintenance 22 6 2 Functional checks 22 6 3 Error analysis 22 6 4 Repairs and service 23 7 Accessories 24 7 1 Required accessories 24 7 2 Preparations 24 8 T...

Страница 6: ...eir oscillating tank bottoms they contain piezoelectric transducers the energy of which is transferred to the bath liquid with ultrasound frequency as mechanical oscillations As a result microscopical...

Страница 7: ...ter machine treatment In this connection they are to be used together with suitable non fixative disinfection and or detergent preparations in order to support or expedite their effect Pursuant to sec...

Страница 8: ...o dewing Only for indoor operation Bath type Order No Interior dimensions L W D Operat ing volume Outlet valve Ultrasonic peak power Ultra sonic nominal output Weight net Heating power Current consump...

Страница 9: ...r use as a medical device Name Ultrasonic bath UMDNS nomenclature ECRI DIMDI 14 263 Purpose See chapter 1 2 Classification Medical Device Directive 93 42 EWG Appendix IX Class I active non invasive no...

Страница 10: ...to be obtained from the implant manufacturer Operation Observe ambient and set up conditions see chapter 1 4 Only plug in the ultrasonic bath to an outlet with a grounded socket Do not operate the ul...

Страница 11: ...tions 1 Instruction manual Additional accessories according to order see delivery note 2 2 Set up assembly Place the ultrasonic bath atop a firm level and dry surface In doing so observe the maximum w...

Страница 12: ...nt 3 1 1 Ultrasound The ultrasound is operated through the turning knob time switch Timed operation Turn knob to the right range of time 1 15 minutes Green control light flashes Once the time has elap...

Страница 13: ...s The heating system works independently from the ultrasound The heating will automatically turn on every time that the bath temperature drops below the set temperature The heat automatically switches...

Страница 14: ...50 C At least drinking quality water must be used to fill the oscillating tank Water without additives is not suited for sonication BANDELIN recommends the TICKOPUR or STAMMOPUR preparations Only use...

Страница 15: ...n decreases however 3 In order to save time during use the bath fluid may be preheated during degassing For an optimum bath temperature observe the specifications of the specimen manufacturer To prote...

Страница 16: ...ontact liquid Step 2 Degassing the fluid Freshly filled bath fluid or fluid that has remained in the oscillating tank for a longer period of time must be degassed prior to use See also chapter 4 4 1 R...

Страница 17: ...sket or inset beaker reduces the ultrasound effect the ultrasound is absorbed Place the more heavily soiled side facing downward Parts with joints are to be fully opened before placing inside Delicate...

Страница 18: ...the service life see chapter 4 4 2 of the bath fluid Heed the specifications of the preparation manufacturer If necessary empty the oscillating tank Step 7 Empty the oscillating tank Layers of contami...

Страница 19: ...s to be heeded The cleaning solution is to be immediately replaced if contamination is evident It is to be replaced however at least every work day Prepare instruments appropriately e g by rinsing the...

Страница 20: ...All aqueous preparations made by DR H STAMM GmbH are prepared pursuant to the regulations of the German Washing and Cleansing Agents Act are biodegradable and as working solutions may be disposed of i...

Страница 21: ...assive protective layer of the stainless steel The stainless steel is activated in this process and it begins to rust The extraneous rust produces localised corrosion of the stainless steel For this r...

Страница 22: ...uested from the manufacturer Customary aluminium foil is used to conduct the test Next a comparison is made with previously generated foils For more detailed information please see the appendix A meas...

Страница 23: ...sse 3 4 12207 Berlin Repair service E mail Tel 49 30 768 80 13 info bandelin com Fax 49 30 768 802 00 13 In the case of returns the general terms and conditions for delivery and payment of BANDELIN el...

Страница 24: ...the cavitation field and do not damage the tank bottom 7 2 Preparations Special preparations are required for use of the ultrasound that are ultrasound compatible i e they are cavitation conducive bi...

Страница 25: ...rasonic bath no longer works it must be disposed of appropriately BANDELIN electronic GmbH Co KG Heinrichstra e 3 4 12207 Berlin Deutschland www bandelin com info bandelin com 49 30 768 80 0 49 30 773...

Страница 26: ...tion water is discharged in the oscillating tank Sound absorbing Inset basket K EM made of stainless steel an alternative to DIN sieve trays in the medical field KT basket holder required Basket holde...

Страница 27: ...156 K 6 L D 156 RK 156 BH K 6 BL D 156 RK 170 H K 7 D 170 RK 255 H K 5 C D 255 D 5 T KW 5 RK 510 H K 10 GH 10 D 510 D 10 T KW 10 0 RK 512 H K 10 B D 510 D 10 T RK 514 H K 14 D 514 K 14 EM KT 14 D 14 T...

Страница 28: ...made of stainless steel for hanging of inset beakers Positioning for optimum utilisation of ultrasound energy Beaker holder ES made of stainless steel to hold 4 inset beakers in larger ultrasonic bath...

Страница 29: ...0 PD 04 SD 06 SD 09 PD 06 EB 05 DE 156 RK 156 BH KD 0 PD 04 SD 06 SD 09 PD 06 EB 05 DE 156 RK 170 H RK 255 H KD 0 PD 04 SD 06 SD 09 PD 06 EB 05 DE 255 K 5 P RK 510 H KD 0 PD 04 SD 06 SD 09 PD 06 EB 0...

Страница 30: ...test tube contents remain visible Tabletting punch holder TH made of stainless steel Holds tabletting punches of diverse diameters Sieve holder SH 7 made of stainless steel For cleaning of single sie...

Страница 31: ...GV 3 SM 6 FE 12 RK 156 BH EK 10 EK 25 EK 50 EK 100 EK 250 GV 3 SM 6 FE 12 RK 170 H FE 12 RK 255 H EK 10 EK 25 EK 50 EK 100 EK 250 GV 3 SM 5 FE 12 RK 510 H EK 10 EK 25 EK 50 EK 100 EK 250 GV 10 RK 512...

Страница 32: ...ils to be compared at any time The perfo rated areas of all foils should have approx the same dimensions and distribution the results are never identical A process validation e g for the treatment of...

Страница 33: ...the manu facturer Type for interior tank dimensions in mm L x W x H Order No for from to FT 1 190 85 60 3190 DT 31 H DT 52 H RK 31 H RK 52 H FT 4 240 140 80 300 150 150 3074 DL 102 H DL 255 H DT 100 H...

Страница 34: ......

Страница 35: ...l 23 4 l 2 6 l RK 255 H 3 8 l 3 7 l 40 ml 3 7 l 80 ml 3 6 l 120 ml 3 6 l 190 ml 3 4 l 380 ml RK 510 H 6 6 l 6 5 l 70 ml 6 4 l 140 ml 6 4 l 200 ml 6 2 l 330 ml 5 9 l 660 ml RK 512 H 8 7 l 8 6 l 90 ml...

Страница 36: ...Note These and further languages of this user instructions as well as further information are to be found on the enclosed CD...

Отзывы: