Customizing the Analyzer Software
5-9
Figure 5-13 Selecting baud rate
Selecting
parity
To choose the “
Odd
”, “
Even
” or “
None
” (default) check,
SELECT
the desired check
from
the
“
Parity
”
drop-down list
, as Figure 5-14 shows.
Figure 5-14 Setting parity
Activating/deactivating
handshake
If the “
Handshake
” function is activated, to start the transmission this analyzer will send a
handshake signal to an external computer and wait for the response. If the computer does not
respond, this analyzer will abort the transmission and give an alarm for the transmission error.
If the “
Handshake
” function is deactivated, this analyzer will transmit data to the external
computer regardless of the response. This function is deactivated by default.
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...