Operating Your Analyzer
6-19
6.8 Running Prediluted Samples
Press [MENU] and
SELECT
”
Count
” to enter the ”
Count
” screen, as Figure 6-12 shows.
Figure 6-12 “Count” screen in prediluted mode
Press [MODE] to select prediluted mode (any mode preceded by
Prediluted
).
z
Select proper reference range as instructed in Chapter 5.3.6 before running
the samples. Otherwise, the obtained results may be erroneously flagged.
6.8.1 Entering Sample Information
You may enter sample information in any of the three modes,
ID only, All info
and
Batch
edit
, depending on which of the three configurations your analyzer has:
ID
only
All
Info
ID only and Batch edit.
ID only
To enter the sample ID of the next sample, you may
At the
“Count
” screen, use the bar-code scanner (if available) to scan the sample ID into the
analyzer; or
At the “
Count
” screen, press [F1] to enter the “
ID
” window and
ENTER
the sample ID.
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...