Operating Your Analyzer
6-6
6.6 Sample Collection and Handling
z
Samples, controls, calibrator and waste are potentially infectious. Wear
proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow
safe laboratory procedures when handling them in the laboratory.
z
Avoid direct contact with blood samples.
z
Do not re-use disposable products.
z
Use clean K
2
EDTA anticoagulant collection tubes, fused silica glass/plastic
test tubes and 20µL borosilicate
glass capillary tubes.
6.6.1 Whole Blood Samples
Collect and handle the whole blood sample as follows:
1. Collect venous blood with a K
2
EDTA (1.5 to 2.2mg/mL ) anticoagulant collection tube;
2. Rapidly and thoroughly mix the blood with the anticoagulant.
z
You should collect at least 2mL venous blood sample.
z
For the whole blood samples to be used for WBC differential or PLT count,
you shall store them at the room temperature and run them within 4 hours
after collection.
z
If you do not need the PLT, MCV and WBC differential results, you can store
the samples in a refrigerator (2
℃
to 8
℃
) for 24 hours. You need to warm the
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...