Using the Calibration Programs
9-11
SELECT
“
Calibration
→
Fresh Blood
” (Figure 9-10) to enter the “
Fresh Blood
” screen
(Figure 9-11).
Figure 9-11
“
Fresh Blood
”
screen
Complete the fresh blood calibration as instructed below:
Determination of the reference values
Reference values may be determined using three or more normal blood samples prepared for
calibration from the reference method or from a reliably calibrated hematology analyzer.
Editing calibration settings
Selecting
count
mode
At “
Fresh Blood
” screen, press [F1] to select the “
Whole Blood
” or “
Prediluted
” mode.
Selecting
sample
When the mode is selected, press [F3] to choose the fresh blood sample whose reference
values you want to set.
Editing the reference value
Press [F2] to enable the edit boxes in the “
Mean
” column, as Figure 9-12 shows.
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...