Using the QC Programs
8-2
8.2 QC With Controls
8.2.1 QC Editing
Entering the “Controls” screen
Press [MENN] to enter the system menu.
SELECT
“
Controls
” (Figure 8-1), to enter the
“
Controls
” screen, as Figure 8-2 shows. At the QC screen you can include a maximum of 12
parameters, WBC, RBC, HGB, PLT, HCT, MCV, MCHC, MCH, Lymph%, Lymph#, Gran% and
Gran# into a QC run.
Figure 8-1 System menu
Figure 8-2
“
Controls” screen
Selecting a QC file
The analyzer provides 9 QC files for you to save QC settings and results. Every QC file can
save the results of a maximum of 31 QC runs. When the saved QC results have reached the
maximum number, the newest result will overwrite the oldest. You can press [F1] to switch the
QC files and the number will be given on the upper left of the screen. After selecting the QC
file, press [F2] to select the “
Whole Blood
” or “
Prediluted
” mode.
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...