Troubleshooting Your Analyzer
11-3
11.3 Errors indicated by error messages
The analyzer can provide error messages. See the tables below for the error messages and
their probable causes and recommended actions. If the problem still remains after you have
tried the recommended solutions, contact Mindray customer service department or your local
distributor.
Error Message
Possible Cause(s)
Recommended Action
Ambient temp.
abnormal
Abnormal ambient
temperature or temperature
transducer error.
1. Enter the “
Service
→
Status
” screen to
check the ambient temperature;
2. If the actual ambient exceeds 15
℃
to
30
℃
, adjust the temperature.
Otherwise, the analysis results may be
unreliable;
3. If the actual temperature is within the
pre-defined range and the problem
remains, contact Mindray customer
service department or your distributor.
Blank
abnormal
1. Contaminated
diluent,
diluent lines or bath;
2. Expired
diluent.
1. Check if the diluent is contaminated or
expired;
2. Enter the “
Count
” screen and press
the aspirate key to do the startup
procedure;
3. If the problem remains, enter the
“
Service
→
Maintenance
” screen and
do the probe cleanser cleaning
procedure as instructed in
Chapter
10.3.6
. When the procedure is
finished, return to the “
Count
” screen
and do the background check again;
4. If the problem remains, contact
Mindray customer service department
or your local distributor.
HGB error
HGB blank voltage within 0 V
to 3.2 V or 4.9 V to 5 V.
1.
Do the Probe cleanser cleaning
procedure as instructed in
Chapter
10.3.6.
;
2. If the problem remains, adjust the
HGB gain as instructed by
Chapter
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...