Using the QC Programs
8-3
Editing QC settings
If there are saved QC results and settings, you need to delete them first. You can press [F5]
to enter “
QC Table
” screen to delete all the results, see
Chapter 8.2.3
for details.
Entering the “
QC Edit
” screen
Press [F3] at “
Controls
” screen to enter the “
QC Edit
” screen (Figure 8-3).
Figure 8-3 “QC Edit” screen
Entering lot number
ENTER
the lot number of the control to be used into the “
Lot No.
”
box, as Figure 8-4 shows.
Figure 8-4 Entering lot number
Entering expiration date
ENTER
the expiration date of the control to be used into the “
Exp. Date
”
box, as Figure 8-5