Maintaining Your Analyzer
10-18
10.3.11 Drain Tubing
z
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
z
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
z
Do the “Drain Tubing” procedure before relocating the analyzer.
You can perform the “
Drain tubing
” procedure to drain the fluidic system. Follow the steps
given below to do so:
1. Press the appropriate arrow keys ([
←
][
→
] [
↑
][
↓
]) as needed to move the cursor to “
Drain
tubing
”;
2. Remove the diluent, rinse and lyse pickup tubes from the back of the analyzer;
3. Press [ENTER] to start the draining process, as Figure 10-24 shows;
Figure 10-24 Draining the fluidic lines
4. When the draining is done, the screen will display “
Turn off this analyzer
” and you
should turn off the power switch.
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...